Chemistries                 65-6265-9381

AISE sc ISE Ammonium Probe

AISE sc ISE Ammonium Probe
Product #: LXV440.99.10012
Unit Price Contact Hach

Cost-Effective Trending Information

The AISE sc Sensor utilizes ion selective electrode (ISE) technology to provide your plant with high level trending information while saving money by eliminating the need for reagents and sample preparation.

Minimal Maintenance with Simple Cartridge Replacement

The sensor operates within in a Cartrical„¢ cartridge that comes factory calibrated so little maintenance is necessary. Cartridge replacement is simple: unscrew the old cartridge, plug in the new one, and the sensor is ready for measurement.

Simple, Accurate Calibration

Easy to perform, fail-safe calibration corrections compensate for naturally occurring calibration drift in ISE instruments. An advanced menu structure allows you to perform corrections without manual entry of values via Ethernet, SD card or Bluetooth*.

What's included?

Ammonium probe, sensor cartridge, user manual

Specifications

Accuracy: 5% of measured value + 0.2 mg/L (with standard solutions) NH4-N
Cable Length: 10 m / 32.8 fixed cable at sensor
  10 m fixed cable at sensor, available up to 100 m
Calibration Method: Sensorcode (calibrated sensor cartridge) or 2-point in-line value and matrix correction
Controller Compatibility: SC200, SC1000, SC4500. All controllers sold separately
Data Transfer: current output, relays and bus communication via sc controller
Flow: < 4 m/s max.
Includes: Ammonium probe, sensor cartridge, user manual
Installation Angle: 45 ° ± 15 ° (vertical in flow direction)
Length: 320 mm
Lower Limit of Detection (LOD): 0.2 mg/L NH4–N
Material: Cartridge: Stainless steel (1.4571), PVC, POM, ABS, NBR
Measurement method: Potentiometric ion-selective measurement
Measurement Range: 0.2 mg/l NH4-N (with standard solutions)
Measuring Interval: Continuous
Measuring uncertainty: ±5% of the measured value + 0.2 mg/L (with standard solution) NH4-N
Operating temperature range: Air: -20 - 45 °C (-4 - 113 °F)
Parameter: NH4-N
pH Range: 5 - 9 pH
Process Connection: 1 inch thread
Protection Class: IP68
Range: 0 - 1000 mg/L NH4-N
Repeatability: ±5 % of the measured value (with standard solution) NH4-N
Response Time: < 3 min
Sample Pressure: 0.3 bar max.
Sample Temperature: 2 - 40 °C (35.6 - 104 °F)
Sensor Cartridge: With CARTRICAL plus technology: compact housing containing calibrated electrodes for ammonium, potassium, reference system and temperature sensor, all calibrated to each other; typical lifetime six months
Sensor Immersion Depth: 0.3 - 3 m (1.0 - 10 ft)
Storage conditions: 5 to 40 °C (41 to 104 °F) sensor cartridge
Warranty: 12 months
Weight: 2.38 kg
Wetted Materials: Probe: Stainless Steel 1.4571; ASA + PC, silicon; PVC and PU;
Sensor cartridge: PVC; POM; ABS; Stainless Steel 1.4571; NBR
What's included?: Ammonium probe, sensor cartridge, user manual