Chemistries                 65-6265-9381

SC1000 Multi-parameter Universal Controller Display Module with GSM/GPRS (for use outside USA)

SC1000 Multi-parameter Universal Controller Display Module with GSM/GPRS (for use outside USA)
Product #: LXV402.99.01002
Unit Price Contact Hach

More Confidence in Your Instrument’s Performance

Available exclusively on Hach instruments, Prognosys predictive diagnostics uses innovative diagnostic software to read multiple inputs from your instrument and alert you to the instrument's overall performance and upcoming maintenance. An easy-to-read dashboard provides instant indication of measurement reliability and service requirements. Process control has never been easier.

Plug and Play Operation

There’s no complicated wiring or set up procedures with the SC1000 controller. Plug any Hach digital sensor into a Probe Module and it’s ready for use. No special ordering or software configuration is needed.

Communication Options to Fit Any Application Need

The Hach SC1000 controller features state-of-the art Modbus TCP/IP communications protocol for seamless integration into a network of devices that support TCP/IP sockets. Use a standard Ethernet cable or connect wirelessly using GSM/GPRS to communicate with your SCADA, PLC or other network. Available with analog 4-20mA, Modbus 485, Profibus DP, or HART digital communication options.

Expandable and Upgradable

The SC1000 controller can adapt to your needs. Add or change probes without having to change the controller. Plus, with a single Display Module, additional Probe Modules and associated sensors can be added or removed depending on operational needs. Fully upgradable software ensures that this system will not be obsolete. Hach service plans are available.

What's included?

Display module with communication (as appropriate), basic user module

Specifications

Alarm: Low alarm point, low alarm point deadband, high alarm point, high alarm point deadband, off delay, and on delay
Analog Output Functional Mode: PID, high/low phasing, setpoint, deadband, overfeed timer, off delay, on delay
Certifications: European Certifications:
CE per 73/23/EEC and 89/336/EEC
TUV-GS to EN 61010-1
EN 61326 Amd's 1 & 2
Communication: Modbus (RS485): Advanced communications/networking with PLC or SCADA system directly from analyzer
Profibus DP/V1 (certified)
GSM/GPRS Quad-band cellular module (FCC and IC approved, EU and US only)
Ethernet service port, RJ45, 10 MB/s
Dimensions (H x W x D): 315 mm x 242 mm x 150 mm
Display: QVGA, 320 x 240 pixels
Display Viewing Area: 111.4 x 83.5 mm
Enclosure Rating: IP65
GSM option: With GSM (non-US)
Inputs: Up to 12 analogue 0-20 mA, maximum impedance 500 Ohms per probe module. Additional inputs are available with additional probe modules.
Material: Plastic
Material Enclosures: Polycarbonate
Mounting: Surface, panel, and pipe (horizontal and vertical) with optional sun shield
Mounting Configurations: Surface, panel, and pipe (horizontal and vertical) with optional sun shield
Operating temperature range: -20 - 55 °C / 0 - 95% relative humidity, non-condensing
Output: Up to 12 analogue 0/4-20 mA, maximum impedance 500 Ohms per probe module.
Additional analogue outputs with additional probe modules. Optional digital communcations via Modbus (RS485) and Profibus DP/V1.
Outputs: Up to 12 analog 0/4-20 mA, maximum impedance 500 Ohms per probe module.

Additional analog outputs with additional probe modules. Optional digital communcations via Modbus® (RS485) and Profibus® DP/V1.
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 240 V AC, 24 V DC
Relays: Up to four SPDT, user-configurable contacts rated 100 to 230 VAC, 5 Amp resistive maximum per probe module. Additional relays are available with additional probe modules.
Storage conditions: -20 to 70 °C (-4 to 158 °F), 0 to 95% relative humidity, non-condensing
Warranty: 12 months
Weight: Approximately 6.5 kg (depending on configuration)
What's included?: Display module with communication (as appropriate), basic user module