Chemistries                 65-6265-9381

EZ2303 Total Manganese + Manganese Mn(II) Analyser

EZ2303 Total Manganese + Manganese Mn(II) Analyser
Product #: EZ2303.XXXXXXXX
Unit Price Contact Hach

What's included?

EZ2303 Manganese Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Reductor, Buffer, Colour & Mixed Reagent Solution) and 1 x empty 10L Reagent Container with Fittings (for EDTA Solution)

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / UL certified
Cycle Time: 20 min Total Mn (dilution + 5 min)
30 min Total Mn & Mn (II)
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 4 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Colorimetric measurement using formaldoxime method at 450 nm
Number of sample streams: 1, 2, 4 or 8
Parameter: Manganese, total; Mn(II), dissolved
Power: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Precision: Better than 2% full scale range for standard test solutions
Protection Class: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Range: 0.02 - 1 mg/L Mn
  0.02 - 1 mg/L Mn

Optional:
0.004 - 0.1 mg/L
0.01 - 0.25 mg/L
0.01 - 0.5 mg/L
0.16 - 4 mg/L (with internal dilution)
0.32 - 10 mg/L (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 12 months
Weight: 25 kg
What's included?: EZ2303 Manganese Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Reductor, Buffer, Colour & Mixed Reagent Solution) and 1 x empty 10L Reagent Container with Fittings (for EDTA Solution)